Inputs:
9 types of TC's (see below)
Standards: IPTS/68/DIN 43710
CJC rejection: 20:1 (0.05°/°C) typical
External resistance: 100ohms maximum
Resistance thermometer
RTD-2/Pt100 2 wire
Standards: DIN 43760
(100Ω 0°C/138.5Ω 100°C Pt)
Bulb current: 0.2mA maximum
Linear process inputs
mV range: 0 to 50mV
Applicable to all inputs SM = sensor maximum
Calibration accuracy: ±0.25%SM ±1°C
Sampling frequency: input 10Hz, CJC 2 sec.
Common mode rejection: Negligible effect up to 140dB, 240V, 50-60Hz
Series mode rejection: 60dB, 50-60Hz
Temperature coefficient: 150ppm/°C SM
Reference conditions: 22°C ±2°C, rated voltage after 15
minutes settling time.Output devices
SSD/SSd1/SSd2: solid state relay driver: To switch a
Remote SSR 5Vdc +0/-15% 15mA non-isolated
Miniature power relay: form A/SPST contacts (AgCdO) RLY
RLY1: 2A/250VAC resistive load
RLY2: 1A/250VAC resistive load |
General
Display: Upper, 4 Digits, high brightness green LED. 10mm (0.4") high.
Digital range -199 to 9999
Hi-res mode -199.9 to 999.9
LED output indicators - flashing SP1 square, green; SP2 round, red
Keypad: 3 elastomeric buttonsEnvironmental
Humidity: Max 80%
Altitude: up to 2000M
Installation: Categories ll and lll
Pollution: Degree ll
Protection: NEMA 4X, lP66
EMC emission: EN50081-1 FCC Rules 15 subpart J
Class A
EMC immunity: EN50082-2
Ambient: 0-50°C (32-130°F)
Moldings: flame retardant polycarbonate
Weight: 130g (4.2 oz)
Cutout: 1/16 DIN -
45mm x 45mm (1.77" x 1.77") -
Punches
|